Smiths Medical MD, Inc. began a recall of Deltec Cozmo Insulin Pump on . The reason FDA records is "The vibratory alarm motor installed in a small number of Deltec Cozmo Insulin Pumps may function intermittently". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0210-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Deltec Cozmo Insulin Pump
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- The vibratory alarm motor installed in a small number of Deltec Cozmo Insulin Pumps may function intermittently.
- FDA root cause
- Other
- Action
- All customers were contacted and instructed to only use the Beeper alarm function and not the vibrate alarm feature. Patients were instructed to return the affected pump to Smiths Medical MD upon receipt of the replacement pump.
- Quantity in commerce
- 6 pumps to patients and 30 pumps in France
- Distribution
- 371 = Number of Individual Product Manufactured 6 = Number of Product Distributed to Patients 365 = Number of Products Either at Smiths Medical MD or at the Distributor or with a Smiths Medical MD Sales Rep.
- Product codes and lots
- Pump Serial Numbers A20721; A20722; A20723; A20724; A20741; A207742