Z-0062-05
Infusion Dynamics Power Infuser pump. Model number M100B-3A
Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics / initiated 2004-09-21 / Terminated
Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics began a recall of Infusion Dynamics Power Infuser pump. Model number M100B-3A on . The reason FDA records is "Front Panel may be cracked on some units. If cracked fluid can enter the infuser and short the start/stop button. Shorting would cause the infuser to stay in the open position". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0062-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics
- Product
- Infusion Dynamics Power Infuser pump. Model number M100B-3A
- Product code
- FRN Pump, Infusion
- Reason for recall
- Front Panel may be cracked on some units. If cracked fluid can enter the infuser and short the start/stop button. Shorting would cause the infuser to stay in the open position.
- FDA root cause
- Other
- Action
- The recalling firm issued a recall letter dated 9/20/04 to their customers informing them of the problem and the need to return the units.
- Quantity in commerce
- 255 units
- Distribution
- The products were shipped to dealers in TX, MO, and VT; medical facilities in CA, NV, and NY. The product was also shipped to government accounts in PA and Germany.
- Product codes and lots
- Serial numbers 13-05863 through 13-05929 and 13-6005 through 13-06036