Z-0196-05

POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505

Siemens Medical Solutions USA, Inc. / initiated 2004-09-23 / Terminated

Siemens Medical Solutions USA, Inc began a recall of POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505 on . The reason FDA records is "Orbital gears may exhibit excess wear". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0196-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Orbital gears may exhibit excess wear
FDA root cause
Other
Action
The recalling firm issued Update Instructions, AX077/04/S, to Siemen''s Customer Service Representative (CSR) to visit affected sites and check the orbital gears for excessive wear. If the CSR finds gears exhibiting excessive wear the orbital gears will be replaced. If no wear is observed the gears will be lubricated as a preventive measure.
Quantity in commerce
82 units
Distribution
The product was shipped to medical facilities nationwide and government accounts in KS and MI.
Product codes and lots
Serial Numbers for model 1148902: 01000 through 01267, 02001 through 02117. Serial numbers for model4784505: 03001 through 035158, 04001 through 04136, and 05001 through 05098