Z-0196-05
POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505
Siemens Medical Solutions USA, Inc. / initiated 2004-09-23 / Terminated
Siemens Medical Solutions USA, Inc began a recall of POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505 on . The reason FDA records is "Orbital gears may exhibit excess wear". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0196-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Orbital gears may exhibit excess wear
- FDA root cause
- Other
- Action
- The recalling firm issued Update Instructions, AX077/04/S, to Siemen''s Customer Service Representative (CSR) to visit affected sites and check the orbital gears for excessive wear. If the CSR finds gears exhibiting excessive wear the orbital gears will be replaced. If no wear is observed the gears will be lubricated as a preventive measure.
- Quantity in commerce
- 82 units
- Distribution
- The product was shipped to medical facilities nationwide and government accounts in KS and MI.
- Product codes and lots
- Serial Numbers for model 1148902: 01000 through 01267, 02001 through 02117. Serial numbers for model4784505: 03001 through 035158, 04001 through 04136, and 05001 through 05098