Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0025-05

Difco (TM) BBL (TM) Gram Crystal Violet, 250 mL bottles in packs of four, catalog # 212525 and 3.8 L bottles...

Becton Dickinson & Co / initiated 2004-09-29 / Terminated

Becton Dickinson & Co. began a recall of Difco (TM) BBL (TM) Gram Crystal Violet, 250 mL bottles in packs of four, catalog # 212525 and 3.8 L bottles, catalog # 212526, staining reagent on . The reason FDA records is "Reagent for microbiological testing may cause inconsistent staining characteristics and subsequent misidentification of bacteria in patient samples". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0025-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
Difco (TM) BBL (TM) Gram Crystal Violet, 250 mL bottles in packs of four, catalog # 212525 and 3.8 L bottles, catalog # 212526, staining reagent.
Product code
JTS Stains, Microbiologic
Reason for recall
Reagent for microbiological testing may cause inconsistent staining characteristics and subsequent misidentification of bacteria in patient samples.
FDA root cause
Other
Action
The recalling firm notified distributors and direct accounts by fax, certified letter, and e-mail on 9/29/04. The notifications advised that staining characteristics may not be appropriate, and that product should be discarded for replacement. Distributors were additionally requested to provide a consignee list for direct notification by BD. Consignees were requested to return response form.
Distribution
Product was sold to distributors and end-users nationwide and to the firm''s international affiliates.
Product codes and lots
Lots 4223220, Exp 5/28/05 4226886, Exp 5/28/05 4233156, Exp 5/28/05 4237134, Exp 6/30/05 4226882, Exp 5/28/05 4237144, Exp 6/30/05