Z-0161-05
SOMATOM CT System Sensation Cardiac (+ Straton), Computed Tomography X-Ray System, Model Numbers 75 43 106
Siemens Medical Solutions USA, Inc. / initiated 2004-09-28 / Terminated
Siemens Medical Solutions USA, Inc began a recall of SOMATOM CT System Sensation Cardiac (+ Straton), Computed Tomography X-Ray System, Model Numbers 75 43 106 on . The reason FDA records is "Calcium Scoring feature on this devices is not closing properly after patient's exams". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0161-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM CT System Sensation Cardiac (+ Straton), Computed Tomography X-Ray System, Model Numbers 75 43 106
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Calcium Scoring feature on this devices is not closing properly after patient's exams.
- FDA root cause
- Other
- Action
- The recalling firm issued a Customer Safety Advisory with Update Instructions, CT036/04/S and CN048.04/2 to their affected customers. The letter informed the customers of the potential issue and provided instructions to avoid the problem while using this feature. A software upgrade is being developed to corect this issue.
- Quantity in commerce
- 47 units
- Distribution
- The product was shipped to medical facilities nationwide and government account in IL.
- Product codes and lots
- Serial Numbers: 52948, 52950, 52951, 52952, 52953, 52954, 52955, 52956, 52957, 52958, 52959, 52960, 52961, 52962, 52963, 52964, 52965, 52966, 52967, 52968, 52969, 52971, 52972, 52973, 52974, 52975, 52976, 52977, 52978, 52979, 52980, 52981, 52982, 52983, 52984, 52985, 52986, 52988, 52989, 58990, 52991, 52992, 52993, 52994, 52996, 52997, and 58999