Z-0163-05
LEONARDO MM-WS, Computed Tomography X-Ray, Model Number 71 29 534
Siemens Medical Solutions USA, Inc. / initiated 2004-09-28 / Terminated
Siemens Medical Solutions USA, Inc began a recall of LEONARDO MM-WS, Computed Tomography X-Ray, Model Number 71 29 534 on . The reason FDA records is "Calcium Scoring feature on this devices is not closing properly after patient's exams". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0163-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- LEONARDO MM-WS, Computed Tomography X-Ray, Model Number 71 29 534
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Calcium Scoring feature on this devices is not closing properly after patient's exams.
- FDA root cause
- Other
- Action
- The recalling firm issued a Customer Safety Advisory with Update Instructions, CT036/04/S and CN048.04/2 to their affected customers. The letter informed the customers of the potential issue and provided instructions to avoid the problem while using this feature. A software upgrade is being developed to corect this issue.
- Quantity in commerce
- 8 units
- Distribution
- The product was shipped to medical facilities nationwide and government account in IL.
- Product codes and lots
- Serial Numbers: 6026, 6037, 6047, 6106, 6114, 6115, 6116, and 20014
FIRST ARTICLE / WEEKLY