Z-0045-05
Advanced Bionics implantable cochlear stimulator CLARION CII Models AB-5100H-11A AB-5100h-12A
Advanced Bionics LLC / initiated 2004-09-27 / Terminated
Advanced Bionics began a recall of Advanced Bionics implantable cochlear stimulator CLARION CII Models AB-5100H-11A AB-5100h-12A on . The reason FDA records is "All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the device to stop functioning". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0045-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advanced Bionics LLC
- Product
- Advanced Bionics implantable cochlear stimulator CLARION CII Models AB-5100H-11A AB-5100h-12A
- Product code
- MCM Implant, Cochlear
- Reason for recall
- All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the device to stop functioning.
- FDA root cause
- Other
- Action
- Expanded recall was initiated with notification to clinicians on 9/27/2004. Expanded effort includes: notice to patients implanted, notice to clinicians, worldwide return of all unimplanted coclear implants.
- Distribution
- U.S. and Canada and France
- Product codes and lots
- Unimplanted CLARION CII