Z-0046-05

Advanced Bionics implantable cochlear stimulator HiRes 90, Models CI-1400-01 CI-1400-02

Advanced Bionics LLC / initiated 2004-09-27 / Terminated

Advanced Bionics began a recall of Advanced Bionics implantable cochlear stimulator HiRes 90, Models CI-1400-01 CI-1400-02 on . The reason FDA records is "All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the device to stop functioning". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0046-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Bionics LLC
Product
Advanced Bionics implantable cochlear stimulator HiRes 90, Models CI-1400-01 CI-1400-02
Product code
MCM Implant, Cochlear
Reason for recall
All unimplanted CLARION and HiResolution cochlear implants due to the potential presence of moisture in the internal circuitry, which can cause the device to stop functioning.
FDA root cause
Other
Action
Expanded recall was initiated with notification to clinicians on 9/27/2004. Expanded effort includes: notice to patients implanted, notice to clinicians, worldwide return of all unimplanted coclear implants.
Quantity in commerce
46 units
Distribution
U.S. and Canada and France
Product codes and lots
HiRes 90K Implantable Cochlear Stimulator, Model Number CI-1400-01, manufactured prior to December 1, 2003.