Z-0176-05
Alto implantable cardioverter defibrillator
Ela Medical, Inc. / initiated 2004-10-21 / Terminated
ela Medical Llc began a recall of Alto implantable cardioverter defibrillator on . The reason FDA records is "Alto Implantable Defibrillators Models 614 and 615. These models can experience sudden no output, and early end of life characteristics due to metal migration in their controlling hybrids, causing high current drain...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0176-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ela Medical, Inc.
- Product
- Alto implantable cardioverter defibrillator
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Alto Implantable Defibrillators Models 614 and 615. These models can experience sudden no output, and early end of life characteristics due to metal migration in their controlling hybrids, causing high current drain leading to premature battery depletion.
- FDA root cause
- Other
- Action
- The company observed occurrences of premature battery depletion and/or prolonged charge time for some Alto 614 and Alto 615 devices. The firm is requesting physician's assistance to monitor the performance of a limited number of implanted Alto 614 and 615 devices.
- Quantity in commerce
- 434 users recived the product
- Distribution
- Nationwide. Devices were received from the manufacturer in France. After received in the US, No devices were distributed from the US to foreign consignees. 95 physicians received the recalled product. ELA Medical has contacted physicians who follow patients implanted with the devices that are the subject of the correction. It should be noted that in many cases, the physician following the patient is not the physician who implanted the device. Because ELA is requesting assistance to monitor the performance of implanted devices, this notification is directed at the physicians who are following the patients.
- Product codes and lots
- Model Numbers DR model 614, VR model 615