Z-0192-05
APL Valve used on some Fabius GS (Catalog number 4117110), Fabius Tiro (Catalog number 4118350) and Narkomed...
Draeger, Inc. / initiated 2004-09-15 / Terminated
Draeger Medical, Inc. began a recall of APL Valve used on some Fabius GS (Catalog number 4117110), Fabius Tiro (Catalog number 4118350) and Narkomed 6000 (Catalog number 4118070) series Anesthesia... on . The reason FDA records is "rotating knob of the rotary style APL valve can become separated from the assembly". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0192-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- APL Valve used on some Fabius GS (Catalog number 4117110), Fabius Tiro (Catalog number 4118350) and Narkomed 6000 (Catalog number 4118070) series Anesthesia Machines.
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- rotating knob of the rotary style APL valve can become separated from the assembly.
- FDA root cause
- Other
- Action
- The recalling firm issued a letter to their accounts informing them of the problem. The letter also informed the accounts that they would be contacted when the replacement knob was available to schedule the correction.
- Quantity in commerce
- 2321 units
- Distribution
- The product was shipped to medical supply companies, medical facilities nationwide. The product was also shipped to a government accounts in AR, FL, ID, LA, MA, MD, ME, NC, NY, OH, TX, VA, and VT. The product was also shipped to Canada, Germany, Korea, China, Australia, Singapore, Switzerland, and England
- Product codes and lots
- APL Valve part number MK00625
FIRST ARTICLE / WEEKLY