Z-0496-05
Proven Porous CR Femur. Implantable orthopedic device
Globus Medical, Inc. / initiated 2004-09-24 / Terminated / root cause: packaging
Stelkast Co began a recall of Proven Porous CR Femur. Implantable orthopedic device on . The reason FDA records is "Sterility may be compromised. The inner pouch of the implant packaging was sticking out beyond the seal of the outer pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0496-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Globus Medical, Inc.
- Product
- Proven Porous CR Femur. Implantable orthopedic device
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Sterility may be compromised. The inner pouch of the implant packaging was sticking out beyond the seal of the outer pouch
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- The recalling firm telephoned their customers 9/24/04 to inform them of the problem and the need to return the product. As a follow up to the telephone call, the recalling firm sent fax notifications to their customers. The medical supply companies were instructed to contact their customers.
- Quantity in commerce
- 105 units
- Distribution
- The product was shipped to medical supply companies in CA, MO, NC, and WV and sales representatives in AZ , NC and PA.
- Product codes and lots
- Catalog number SC2113-2 Lot number 3763-012501, Catalog number SC2113-3 Lot number 3764-012501, Catalog number SC2113-4 Lot number 3765-012501, Catalog number SC2113-1 Lot number 3762-012501, Catalog number SC2113-6 Lot number 3767-012501, Catalog number SC2113-5 Lot number 3766-012501, Catalog number SC2114-4 Lot number 3771-012501, Catalog number SC2114-3 Lot number 3770-012501, Catalog number SC2114-1 Lot number 3768-012501, Catalog number SC2114-2 Lot number 3769-012501, Catalog number SC2114-6 Lot number 3773-012501, and Catalog number SC2114-5 Lot number 3772-012501.