Z-0175-05
Stryker brand Neptune waste management system, Neptune Rover, 15 amp plug; model 0700-003-000
Stryker Corp. / initiated 2004-10-12 / Terminated
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker brand Neptune waste management system, Neptune Rover, 15 amp plug; model 0700-003-000 on . The reason FDA records is "The grounding pin may break inside of the electrical plug, resulting in a shock hazard". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0175-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker brand Neptune waste management system, Neptune Rover, 15 amp plug; model 0700-003-000.
- Product code
- FYD Apparatus, Exhaust, Surgical
- Reason for recall
- The grounding pin may break inside of the electrical plug, resulting in a shock hazard.
- FDA root cause
- Other
- Action
- Hospital customers will be sent a recall notice dated 11/09/04, new plugs for installation on their units, and instructions for installation.
- Quantity in commerce
- 365
- Distribution
- United States and Canada.
- Product codes and lots
- All units with an Interpower brand AC power plug, which includes serial numbers 0203177 through 0406301403.