Z-0432-05
Harvard 1 Single Channel Syringe Infusion Pump, P/N 2003-001
Harvard Clinical Technology / initiated 2004-11-03 / Terminated
Harvard Clinical Technology began a recall of Harvard 1 Single Channel Syringe Infusion Pump, P/N 2003-001 on . The reason FDA records is "Potential pump motor/encoder assembly failure may cause over infusing medication in patients". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0432-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Harvard Clinical Technology
- Product
- Harvard 1 Single Channel Syringe Infusion Pump, P/N 2003-001
- Product code
- MEA Pump, Infusion, Pca
- Reason for recall
- Potential pump motor/encoder assembly failure may cause over infusing medication in patients
- FDA root cause
- Other
- Action
- Harvard Clinical Technology notified consignees by letter on 11/3/04. Harvard will contact consignees by telephone to schedule a return. The motors inside the recalled products will be replaced at Harvard Clinical. Harvard will provide Loaner pumps upon request during the repair process.
- Quantity in commerce
- 154 units
- Distribution
- AK, KY, TX, MA, NY, CA, GA, OK, IL, CT, LA, FL, MD,NC, NJ, VA, MI, WI, IA, WV,OH,
- Product codes and lots
- Serial number range: 308102012 - 408102213