Z-0209-05

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required)...

Terumo Cardiovascular Systems Corporation / initiated 2003-08-12 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764 on . The reason FDA records is "The central control monitor screen may go blank during use because of a bad wiring connection in the wiring harness or a low voltage setting on the potentiometer". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0209-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The central control monitor screen may go blank during use because of a bad wiring connection in the wiring harness or a low voltage setting on the potentiometer.
FDA root cause
Other
Action
Service representatives were sent a service bulletin dated 8/12/03 instructing them to correct units in the field bearing serial numbers between 20 and 109.
Distribution
United States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Japan, Korea, Singapore and United Kingdom.
Product codes and lots
All units bearing serial numbers 20 through 109.