Z-0261-05

AMAX ThromboMAX with Calcium, 10x4ml vials. Catalog #T9902. Lyophilized extract of rabbit brain with buffer...

Trinity Biotech USA / initiated 2004-10-27 / Terminated

Trinity Biotech USA began a recall of AMAX ThromboMAX with Calcium, 10x4ml vials. Catalog #T9902. Lyophilized extract of rabbit brain with buffer, stabilizers and calcium chloride. Responsible firm... on . The reason FDA records is "Prolongation of Prothrombin times in patient samples, resulting in a falsely elevated INR value". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0261-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Trinity Biotech USA
Product
AMAX ThromboMAX with Calcium, 10x4ml vials. Catalog #T9902. Lyophilized extract of rabbit brain with buffer, stabilizers and calcium chloride. Responsible firm on the label: Trinity Biotech PLC, IDA Business Park, Bray, County Wicklow, Ireland. USA Enquiries - Trinity Biotech USA, 1930 Innerbelt Business Center Drive, St. Louis, MO 63114, USA.
Product code
GJS Test, Time, Prothrombin
Reason for recall
Prolongation of Prothrombin times in patient samples, resulting in a falsely elevated INR value.
FDA root cause
Other
Action
Letters dated 11/1/2004 with instructions to (1) Discard remaining kits; (2) notify end-users; (3) Evaluate patient records when the kits were used to monitor Oral Anti-coagulant Therapy.
Quantity in commerce
1,183 kits
Distribution
Trinity Biotech, Jamestown, NY sold product to 87 customers: 10 distributors (7 in U.S. & 3 foreign); 1 exporter; and 76 end users. 84 of those customers are located in the United States. There is one consignee in Canada, one in Brazil, and one in Venezuela. In addition, the manufacturer in Ireland sold additional units of this lot to other countries.
Product codes and lots
M13906, Exp. 6/14/06