Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0310-05

Brand name: NIM EMG Endotracheal Tube REF 82-29106, 82-29107, 82-29108 NIM EMG Reinforced Endotracheal Tube...

Medtronic Xomed, Inc. / initiated 2004-10-29 / Terminated

Medtronic Xomed, Inc. began a recall of Brand name: NIM EMG Endotracheal Tube REF 82-29106, 82-29107, 82-29108 NIM EMG Reinforced Endotracheal Tube REF 82-29306, 82-29307, 82-29308 [All products are... on . The reason FDA records is "The firm received six reports of tubes with wires 'out of channel' two of which involved injuried patients. In one case in Germany it was reported that the patient had soreness of the throat and was discharged after...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0310-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Xomed, Inc.
Product
Brand name: NIM EMG Endotracheal Tube REF 82-29106, 82-29107, 82-29108 NIM EMG Reinforced Endotracheal Tube REF 82-29306, 82-29307, 82-29308 [All products are packaged one unit per box.]
Product code
ETN Stimulator, Nerve
Reason for recall
The firm received six reports of tubes with wires 'out of channel' two of which involved injuried patients. In one case in Germany it was reported that the patient had soreness of the throat and was discharged after four days. In the next case the patient had a tracheal wound. It was not clear what caused the wound.
FDA root cause
Other
Quantity in commerce
22569 EMG Tubes
Distribution
Distributed nationwide and worldwide. to countries in the five continents with the highest populations.
Product codes and lots
Product Code: 77 ETN  Lot Numbers: 32423300 through 35505400, and 35542200, 35542300, 35542400, 35542500, 35542600, 35543400, 35597200, 35602100, 35633400, 35633600, 35638200, 35638400