Z-0213-05

Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic...

Medtronic Powered Surgical Solutions / initiated 2004-11-03 / Terminated

Power Surgical Solutions began a recall of Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas on . The reason FDA records is "Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke)". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0213-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Powered Surgical Solutions
Product
Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
Product code
HBB Motor, Drill, Pneumatic
Reason for recall
Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke).
FDA root cause
Other
Action
The firm initiated the recall by letter on November 3, 2004 asking consignees to return unused units.
Quantity in commerce
12,286 cases of 4 units per case
Distribution
Product was shipped nationwide and to the following foreign countries: Australia, Canada, Hong Kong, Japan, Netherlands, Columbia, Chile, and Venezuela.
Product codes and lots
A5972-V01; A5982-V01; A5991-V01; A5992-V01; A6308-V01through A6327-V01; A6666-V01 through A6684-V01; A6686-V01 through A6691-V01; A6919-V01 through A6934-V01; A7184-V01 through A7188-V01; A7194-V01through A7198-V01; A7495-V01 through A7501-V01; A7520-V01; A7521-V01; and A7525-V01