Z-0213-05
Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic...
Medtronic Powered Surgical Solutions / initiated 2004-11-03 / Terminated
Power Surgical Solutions began a recall of Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas on . The reason FDA records is "Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0213-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Powered Surgical Solutions
- Product
- Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
- Product code
- HBB Motor, Drill, Pneumatic
- Reason for recall
- Reduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke).
- FDA root cause
- Other
- Action
- The firm initiated the recall by letter on November 3, 2004 asking consignees to return unused units.
- Quantity in commerce
- 12,286 cases of 4 units per case
- Distribution
- Product was shipped nationwide and to the following foreign countries: Australia, Canada, Hong Kong, Japan, Netherlands, Columbia, Chile, and Venezuela.
- Product codes and lots
- A5972-V01; A5982-V01; A5991-V01; A5992-V01; A6308-V01through A6327-V01; A6666-V01 through A6684-V01; A6686-V01 through A6691-V01; A6919-V01 through A6934-V01; A7184-V01 through A7188-V01; A7194-V01through A7198-V01; A7495-V01 through A7501-V01; A7520-V01; A7521-V01; and A7525-V01