Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0569-05

ReliOn Insulin Syringe 0.5cc 30 Gauge 5/16 inch (8mm) Needle (U-100 Insulin). NDC 81306-0515-23

Becton Dickinson & Co / initiated 2004-10-22 / Terminated / root cause: labeling and UDI

Becton Dickinson & Company began a recall of ReliOn Insulin Syringe 0.5cc 30 Gauge 5/16 inch (8mm) Needle (U-100 Insulin). NDC 81306-0515-23 on . The reason FDA records is "Syringes labeled as 5/16 inch product (30 ga) actually contained the 1/2 inch product (28 ga)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0569-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
ReliOn Insulin Syringe 0.5cc 30 Gauge 5/16 inch (8mm) Needle (U-100 Insulin). NDC 81306-0515-23.
Product code
FMF Syringe, Piston
Reason for recall
Syringes labeled as 5/16 inch product (30 ga) actually contained the 1/2 inch product (28 ga).
Root cause tag
labeling and UDI
FDA root cause
Other
Action
BD has sent letters to the distributor McKesson Inc and the retailer Wal-Mart Inc.
Quantity in commerce
1543000 syringes
Distribution
The syringes were sent to McKesson Corp, 120 Senlac Drive, Carrollton, TX 75006 who then sold them to Wal-Mart Corp, 702 SW 8th Street, Bentonville, AR 72716. There are no other customers.
Product codes and lots
Batch # 4083690