Z-0371-05
UniCel Dxl 800 Access Immunoassay Systems (Reaction Vessels)
Siemens Healthcare Diagnostics Inc. / initiated 2004-10-29 / Terminated / root cause: packaging
Beckman Coulter Inc began a recall of UniCel Dxl 800 Access Immunoassay Systems (Reaction Vessels) on . The reason FDA records is "Beckman Coulter has confirmed reports of rare occurrences of UniCel Dxl 800 Access Immunoassay Systems Reaction Vessels that are deformed or flattened in their packaging bags for lot 04250160. Deformed or flattened RVs...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0371-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- UniCel Dxl 800 Access Immunoassay Systems (Reaction Vessels).
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Beckman Coulter has confirmed reports of rare occurrences of UniCel Dxl 800 Access Immunoassay Systems Reaction Vessels that are deformed or flattened in their packaging bags for lot 04250160. Deformed or flattened RVs could present a rish of causing jams on the UniCel Exl 800 Access Immunoassay Systems that may require service intervention and could compromise sample analysis.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- All customers that were shipped lot 042500160 of the reaction vessels were contacted via telephone (or via on-site visit - Canada only) October 29 - November 1, 2004. A Product Correction Action Letter dated November 2,2 004, sent via U.S. mail to all customers, once contacted by phone with fax back forms.
- Quantity in commerce
- 56 boxes
- Distribution
- FL, HI, IL, MS, MI, NC, NY, OH, OK, TX, UT, VA, WI, and Canada
- Product codes and lots
- Part Number: 386167 Lot Number: 04250160