Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0374-05

Warm Air Model 135 Warming Units

Gentherm Medical, LLC / initiated 2004-11-05 / Terminated

Cincinnati Sub-Zero Products began a recall of Warm Air Model 135 Warming Units on . The reason FDA records is "Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0374-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gentherm Medical, LLC
Product
Warm Air Model 135 Warming Units.
Product code
BTE
Reason for recall
Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device.
FDA root cause
Other
Action
The firm notified consignees by certified letter on November 5, 2004.
Quantity in commerce
649 units.
Distribution
The product was distributed throughout the United States and to foreign countries including Saudi Arabia, Taiwan, Japan, and Canada.
Product codes and lots
All serial numbers.