Cincinnati Sub-Zero Products began a recall of Warm Air Model 135 Warming Units on . The reason FDA records is "Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0374-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gentherm Medical, LLC
- Product
- Warm Air Model 135 Warming Units.
- Product code
- BTE
- Reason for recall
- Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device.
- FDA root cause
- Other
- Action
- The firm notified consignees by certified letter on November 5, 2004.
- Quantity in commerce
- 649 units.
- Distribution
- The product was distributed throughout the United States and to foreign countries including Saudi Arabia, Taiwan, Japan, and Canada.
- Product codes and lots
- All serial numbers.