Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer
Beckman Coulter Inc began a recall of Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer on . The reason FDA records is "A rare condition of the Access 2 Immunoassay Systems software versions 2.0, 2.1 and 2.2 could present a risk of an erroneous confirmatory result under a specific set of circumstances. The Chlamydia Blocking results may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0265-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Access Immunoassay Systems, 862.2160 discrete photometric chemistry analyzer.
- Product code
- CGN Acid, Folic, Radioimmunoassay
- Reason for recall
- A rare condition of the Access 2 Immunoassay Systems software versions 2.0, 2.1 and 2.2 could present a risk of an erroneous confirmatory result under a specific set of circumstances. The Chlamydia Blocking results may be affected. The instrument will erroneously perform the blocking calculation on the affected test by using the mean of the diluted and neat results.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Product Corrective Action (PCA) letter that informs customers (who ordered Access Chlamydia reagent in the past 24 months and have an Access 2 Immunoassay System) of this potential issue sent via US mail on November 8, 2004. The letter instructs customers to use unique Sample IDs for all samples assayed on the system - including those which have been manipulated offline prior to assay. A Technical Information Update will be issued to inform all support personnel of this issue.
- Quantity in commerce
- 1,197
- Distribution
- Nationwide and Canada
- Product codes and lots
- Access 2 Immunoassay Systems Part Number 81600N software versions 20., 2.1, and 2.2. Access Immunoassay Systems Chlamydia Reagent Kit Part Number 34401.