ACCU-CHEK Inform meter; catalog No. 3035123

Roche Diagnostics GmbH / initiated 2004-11-19 / Terminated / root cause: software

Roche Diagnostics Corp. began a recall of ACCU-CHEK Inform meter; catalog No. 3035123 on . The reason FDA records is "A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0372-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Inform meter; catalog No. 3035123.
Product code
NBW System, Test, Blood Glucose, Over The Counter
Reason for recall
A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa.
Root cause tag
software
FDA root cause
Other
Action
Consignees were notified via a letter dated 11/19/04, which instructed them to discontinue using the OTE functionality.
Quantity in commerce
4697
Distribution
Nationwide.
Product codes and lots
All units with firmware version 2.3 that have the OTE (other test entry) functionality enabled.