Z-0372-05
ACCU-CHEK Inform meter; catalog No. 3035123
Roche Diagnostics GmbH / initiated 2004-11-19 / Terminated / root cause: software
Roche Diagnostics Corp. began a recall of ACCU-CHEK Inform meter; catalog No. 3035123 on . The reason FDA records is "A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0372-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- ACCU-CHEK Inform meter; catalog No. 3035123.
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Consignees were notified via a letter dated 11/19/04, which instructed them to discontinue using the OTE functionality.
- Quantity in commerce
- 4697
- Distribution
- Nationwide.
- Product codes and lots
- All units with firmware version 2.3 that have the OTE (other test entry) functionality enabled.