Z-0307-05
Easy Go Vac Aspirator, Model PM65. Portable suction unit
Precision Medical, Inc. / initiated 2004-11-09 / Terminated
Precision Medical, Inc. began a recall of Easy Go Vac Aspirator, Model PM65. Portable suction unit on . The reason FDA records is "Battery charger can be overcharged and overheat". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0307-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Precision Medical, Inc.
- Product
- Easy Go Vac Aspirator, Model PM65. Portable suction unit
- Product code
- BTA Pump, Portable, Aspiration (Manual Or Powered)
- Reason for recall
- Battery charger can be overcharged and overheat
- FDA root cause
- Other
- Action
- The recalling firm sent Recall letters via first class mail on 11/15/04 to their consignees. The recalling firm will replace the chargers free of charge.
- Quantity in commerce
- 6000 units
- Distribution
- The product was shipped to medical supply companies and medical facilities nationwide. The product was also shipped to Saudi Arabia, Boliva, Hong Kong, Israel, Brazil, Phillipines, Poland, Canada, Ireland, New Zealand, France, United Kingdom, and Singapore.
- Product codes and lots
- Serial Numbers:008360 to 14360 Charger part number 503974 (Globtek Part # GTM21097CC-5015-1.3 Part No: GS-307-B)