Z-0383-05

Roche brand Cardiac Reader

Roche Diagnostics GmbH / initiated 2004-12-03 / Terminated

Roche Diagnostics Corp. began a recall of Roche brand Cardiac Reader on . The reason FDA records is "Potential for false negative Troponin T or falsely decreased Troponin T and Myoglobin results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0383-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche brand Cardiac Reader.
Reason for recall
Potential for false negative Troponin T or falsely decreased Troponin T and Myoglobin results.
FDA root cause
Other
Action
User hospitals were notified via phone, letter and fax (beginning on 12/3/04) a copy of the recall notification on December 3, 2004, which instructed them to immediately cease using the device. The firm issued a press release on December 8, 2004.
Quantity in commerce
95
Distribution
United States.
Product codes and lots
All units.