Hospitak, REF 962-E, Adapter 22mm/15mm, manufactured by Unomedical, Inc of McAllen, Texas and assembled in...

Unomedical A/S / initiated 2004-12-02 / Terminated

Unomedical, Inc began a recall of Hospitak, REF 962-E, Adapter 22mm/15mm, manufactured by Unomedical, Inc of McAllen, Texas and assembled in Mexico on . The reason FDA records is "Adapters have been found to be blocked or occluded - potentially preventing exhalation or inhalation". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0285-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Unomedical A/S
Product
Hospitak, REF 962-E, Adapter 22mm/15mm, manufactured by Unomedical, Inc of McAllen, Texas and assembled in Mexico
Product code
BZA Connector, Airway (Extension)
Reason for recall
Adapters have been found to be blocked or occluded - potentially preventing exhalation or inhalation
FDA root cause
Other
Action
The firm issued a Public Safety Alert on 11/30/2004. The firm expanded their safety alert on 12/02/2004 and issued press informing of a recall. The firm sent recall letters to consignees on 12/02/2004.
Quantity in commerce
979 cases; 50 units/case (48,950 units total)
Distribution
The product was distributed nationwide to hospitals, home healthcare, and medical transport. Distribution to foreign countries included: Canada; Russia; France; Switzerland; Argentina; Uruguay; and Mexico
Product codes and lots
Lot numbers 04-37, 04-38, 04-40, 04-41, 04-44, 04-45, 04-46, 04-47 and 04-48