Z-0387-05

Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90...

Ortho-Clinical Diagnostics, Inc. / initiated 2004-11-22 / Terminated

Ortho-Clinical Diagnostics began a recall of Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics... on . The reason FDA records is "Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0387-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.
Product code
DCK C-Reactive Protein, Antigen, Antiserum, And Control
Reason for recall
Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
FDA root cause
Other
Action
Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
Quantity in commerce
Cat. 1926740 - 20,903 boxes; Cat 8097990 - 34,877 boxes
Distribution
Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
Product codes and lots
Coatings 401 thru 419.