Z-0387-05
Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90...
Ortho-Clinical Diagnostics, Inc. / initiated 2004-11-22 / Terminated
Ortho-Clinical Diagnostics began a recall of Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics... on . The reason FDA records is "Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0387-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.
- Product code
- DCK C-Reactive Protein, Antigen, Antiserum, And Control
- Reason for recall
- Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
- FDA root cause
- Other
- Action
- Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
- Quantity in commerce
- Cat. 1926740 - 20,903 boxes; Cat 8097990 - 34,877 boxes
- Distribution
- Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
- Product codes and lots
- Coatings 401 thru 419.