Z-0428-05

Pneumotach, Disposable (B) Catalog Number: 4408 Labeled in part: SINGLE USE CATALOG NO.4408 PNEUMOTACH...

Bio-Med Devices, Inc. / initiated 2004-12-06 / Terminated

Bio-Med Devices, Inc. began a recall of Pneumotach, Disposable (B) Catalog Number: 4408 Labeled in part: SINGLE USE CATALOG NO.4408 PNEUMOTACH, DISPOSABLE (B) =-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=... on . The reason FDA records is "Adapters may be occluded potentially preventing inhalation". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0428-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Med Devices, Inc.
Product
Pneumotach, Disposable (B) Catalog Number: 4408 Labeled in part: SINGLE USE CATALOG NO.4408 PNEUMOTACH, DISPOSABLE (B) =-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-=-= CONTENTS: 1- PNEUMOTACH HEAD ASSEMBLY 1- PARATUBE 60'' 2- TUBING FITTINGS 1- 22x22mm ADAPTER
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Adapters may be occluded potentially preventing inhalation
FDA root cause
Other
Action
Bio-Med Devices issued recall notifcation by fax on 12/7/04. Accounts are requested to discontine use and return product. A Press Release issued to AP 12/8/04
Quantity in commerce
107 cs
Distribution
Nationwide Forign: Canada, Greece, Japna, Israel, Saudi Arabia,
Product codes and lots
Lot Numbers: 05092204, 05101104, 05101304, 05101404 05102004, 05102604, 05102804, 05110204 05110304, 05110904, 05111804, 05112404 05120104