Z-0341-05
Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)
Philips North America LLC / initiated 2004-12-15 / Terminated / root cause: software
Philips Medical Systems began a recall of Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use) on . The reason FDA records is "In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0341-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Philips notified customers on 12/13/04 by letter. Users are provided with action required to Mitigate Risk' until the correction is made by a Systems representative.
- Quantity in commerce
- 2520 units
- Distribution
- Nationwide Foreign:Canada, Austria, Australlia, Belgium, China, Denmark, Finland, France, Germany, Israel, Italy,Kuwait, Latvia, Lebannon, Libya, Netherland,Norway, Portugal, Qatar, Russia, Singapore,South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand,Tunisia, UK, United Arab Emirates, Yemen
- Product codes and lots
- Software versions A.02.00 and lower
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
