Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0341-05

Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)

Philips North America LLC / initiated 2004-12-15 / Terminated / root cause: software

Philips Medical Systems began a recall of Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use) on . The reason FDA records is "In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0341-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy
Root cause tag
software
FDA root cause
Other
Action
Philips notified customers on 12/13/04 by letter. Users are provided with action required to Mitigate Risk' until the correction is made by a Systems representative.
Quantity in commerce
2520 units
Distribution
Nationwide Foreign:Canada, Austria, Australlia, Belgium, China, Denmark, Finland, France, Germany, Israel, Italy,Kuwait, Latvia, Lebannon, Libya, Netherland,Norway, Portugal, Qatar, Russia, Singapore,South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand,Tunisia, UK, United Arab Emirates, Yemen
Product codes and lots
Software versions A.02.00 and lower