Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric...
Smiths Medical MD, Inc. began a recall of Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for... on . The reason FDA records is "Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0468-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for this same Burette Administration Set described as 110 inch length/279 cm, 16.5ml priming volume, 60 drops/ml, 3 standard injection sites. Model Number I8C1630 (Reorder No. 21-0319-25) for product distributed Internationally. Model Number 8C1630 (Reorder No. 21-0306-01) for product distributed within the United States.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can allow fluid to leask past and may cause an inaccurate infusion rate..
- FDA root cause
- Other
- Action
- Customers were sent a product Safety & Recall notification via email or regular mail begining 12/03/04 through 12/08/04. A Product Recall form requested the customer for feedback as to Quantity Used, Quantity to be Retained and Quantity to be returned . The Customer Recall Notication states that if no Administration Sets from alternative lots are available and the use of the volumetric pump is essential, follow the instructions in the attached Customer Information Bulletin, which describes how to identify a set which does not have the defect. Customers who find product subject to the recall are to contact Smith Medical Customer Service Department to make arrangements to have the affected products returned to Smiths.
- Quantity in commerce
- 6,690 units total: 3,600 units (120 cases) Internationally & 3,090 units (103 cases) within USA. Each case contains 30 units.
- Distribution
- worldwide.
- Product codes and lots
- Model # I8C1630 (Reorder No. 21-0319-25) distributed outside the United States contains Lot No.: APR04T02. Model # 8C1630 (Reorder No. 21-0306-01) distributed within the United States contains Lot No.: MAR04T04.