Z-0473-05

Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric...

Abbott Medical / initiated 2004-12-03 / Terminated

Smiths Medical MD, Inc. began a recall of Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for... on . The reason FDA records is "Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0473-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric Infusion Pump Systems. 2 Model Numbers are used for this same Primary Checkvalve Set for secondary (piggyback) Administration described as 105 inch length/267 cm, 25ml priming volume, 20 drops/ml, 2 standard injection sites. Model Number I8C4220 (Reorder No. 21-0349-25) for product distributed Internationally. Model Number 8C4220 (Reorder No. 21-0328-01) for product distributed within the United States.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can allow fluid to leask past and may cause an inaccurate infusion rate..
FDA root cause
Other
Action
Customers were sent a product Safety & Recall notification via email or regular mail begining 12/03/04 through 12/08/04. A Product Recall form requested the customer for feedback as to Quantity Used, Quantity to be Retained and Quantity to be returned . The Customer Recall Notication states that if no Administration Sets from alternative lots are available and the use of the volumetric pump is essential, follow the instructions in the attached Customer Information Bulletin, which describes how to identify a set which does not have the defect. Customers who find product subject to the recall are to contact Smith Medical Customer Service Department to make arrangements to have the affected products returned to Smiths.
Quantity in commerce
53,120 units total: 24,000 units (600 cases) Internationally & 29,120 units (728 cases) within USA. Each case contains 40 units.
Distribution
worldwide.
Product codes and lots
Model # I8C4220 (Reorder No. 21-0349-25) distributed outside the United States contains Lot No. FEB04T24 & MAR04T14. Model # 8C4220 (Reorder No. 21-0328-01) distributed within the United States contains Lot No.: MAR04T21 & APR04T08.