Z-0493-05

VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US...

Ethicon, Inc / initiated 2004-12-16 / Terminated / root cause: manufacturing process

Ethicon, Inc began a recall of VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/Manufactured for GYNECARE, A division of Ethicon, Inc... on . The reason FDA records is "Deformation of the Verspoint Hysteroscopic System Connector Cable hand grips when autoclaved causing possible non-sterile electrodes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0493-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon, Inc
Product
VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/Manufactured for GYNECARE, A division of Ethicon, Inc., a Johnson & Johnson Company, Somerville, NJ 08876. The product subject of this recall is the connector cable which goes between the electrosurgical generator and the elctrode used for surgery. Specifically, the recall focuses on the hand grip on the electrode, formed by overmolding a polymer onto the cable, which deforms upon sterilization.
Product code
HIH Hysteroscope (And Accessories)
Reason for recall
Deformation of the Verspoint Hysteroscopic System Connector Cable hand grips when autoclaved causing possible non-sterile electrodes.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recall letter was mailed to the Director of Material Management and the Director of Operating Room on December 14, 2004.
Quantity in commerce
891electrodes
Distribution
Ethicon distributed the devices to 166 hospitals in the US. There are no Government sales. Product sold to other countries was distributed directly from the manufacturerin England to Ethicon distributors outside the US. The Foreign countries with accounts are: Austria, Czech Republic, Greece, Italy, Benelux, Poland, Portugal, Slovenia, Spain, Sweden, France, Germany, Ireland, U.K. Canada, China, Mexico, Taiwan, Venezuela, Korea, Australia, Brazil, and India.
Product codes and lots
Product Code: 00480; Lot #s:  0404086, 0404121, 0404152, 0405010, 0405123, 0406014, 0406068, 0406130, 0407104, 0408031, 0409029, 0409080, 0410021, 0410103, 0411040.