Logix-CM Compounder Software, catalog 2M8400; Baxter Healthcare Corporation, Medication Delivery System...

Baxter Healthcare Corporation / initiated 2004-12-20 / Terminated / root cause: software

Baxter Healthcare Corp. began a recall of Logix-CM Compounder Software, catalog 2M8400; Baxter Healthcare Corporation, Medication Delivery System, Round Lake, IL 60073 on . The reason FDA records is "Logix-CM software version 3.0.4 has not completed its official release process". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0394-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Logix-CM Compounder Software, catalog 2M8400; Baxter Healthcare Corporation, Medication Delivery System, Round Lake, IL 60073
Product code
LHI Set, I.V. Fluid Transfer
Reason for recall
Logix-CM software version 3.0.4 has not completed its official release process.
Root cause tag
software
FDA root cause
Other
Action
Urgent Product Recall letters dated 12/20/04 were sent to the end users via overnight mail. The accounts were informed that software version 3.0.4 has not been officially released and must be removed from their compounders. They were requested to complete and fax the customer reply form to Baxter at 847-270-5457, indicating if they are currently using Logix-CM software. Upon receipt of the reply form, Baxter will schedule a visit to the facility to remove Logix-CM software version 3.0.4 from their compounder and revert the software back to the configuration prior to Logix-CM 3.0.4 placement. Any questions were directed to Baxter Clintec Nutrition, option 1, at 1-800-422-2751.
Quantity in commerce
5 units
Distribution
Arizona, Utah, North Carolina, Wisconsin and California.
Product codes and lots
Software version 3.0.4