Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0725-05

Endopath 5 mm Babcocks with Ratchet Handles (product code 5BB)

Ethicon Endo-Surgery Inc / initiated 2004-12-15 / Terminated

Ethicon Endo-Surgery began a recall of Endopath 5 mm Babcocks with Ratchet Handles (product code 5BB) on . The reason FDA records is "Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0725-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon Endo-Surgery Inc
Product
Endopath 5 mm Babcocks with Ratchet Handles (product code 5BB).
Product code
MDM Instrument, Manual, Surgical, General Use
Reason for recall
Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.
FDA root cause
Other
Action
The recalling firm sent letters to hospitals and distributors, dated 12/14/04.
Quantity in commerce
18,880 devices.
Distribution
The devices were distributed to hospitals and distributors located throughout the United States and to foreign consignees located in Saudi Arabia, Austria, Egypt, United Kingdom, Germany, France, Lebanon, The Netherlands, Spain, Pakistan, South Africa, Italy, United Arab Emirate, Czech Republic, Sweden, Switzerland, Greece, India, Portugal, Slovenia, Hungary, Palestine, and Kuwait.
Product codes and lots
V41T4F, V41W66, V41Y2L, V41Z4A, V4230C, V42553, V4272X, V42A8K, V42E42, V42G9Y, V42M4A, V42M4C, V42P7P, V42T67, V42V3N, V42X38, V4308K, V4333L and samples.