Z-0536-05
COULTER LH 750 Hematology Analyzer
Siemens Healthcare Diagnostics Inc. / initiated 2004-12-16 / Terminated
Beckman Coulter Inc began a recall of COULTER LH 750 Hematology Analyzer on . The reason FDA records is "A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0536-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- COULTER LH 750 Hematology Analyzer.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.
- FDA root cause
- Other
- Action
- Letter 12/20/04 A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation''s database are not affected. If the LH750 instrument displays a pop-up winder titled Database Error (XXXXXX) and text similar to (Microsoft) (ODBC SQL...) As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Do not acknowledge the error. Instead, restart the entire system by completely powering it OFF and then ON. Or, if the LH750 Workstation experiences a change in performance, such as screen display freezes, slow response time, or requires frequent resets to continue processing. As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Verify the accuracy of all patient reports for the samples until service confirms the performance of system.
- Quantity in commerce
- 1,257
- Distribution
- US and Canada
- Product codes and lots
- Part Number 6605632