Z-0536-05

COULTER LH 750 Hematology Analyzer

Siemens Healthcare Diagnostics Inc. / initiated 2004-12-16 / Terminated

Beckman Coulter Inc began a recall of COULTER LH 750 Hematology Analyzer on . The reason FDA records is "A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0536-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
COULTER LH 750 Hematology Analyzer.
Product code
GKZ Counter, Differential Cell
Reason for recall
A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.
FDA root cause
Other
Action
Letter 12/20/04 A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation''s database are not affected. If the LH750 instrument displays a pop-up winder titled Database Error (XXXXXX) and text similar to (Microsoft) (ODBC SQL...) As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Do not acknowledge the error. Instead, restart the entire system by completely powering it OFF and then ON. Or, if the LH750 Workstation experiences a change in performance, such as screen display freezes, slow response time, or requires frequent resets to continue processing. As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Verify the accuracy of all patient reports for the samples until service confirms the performance of system.
Quantity in commerce
1,257
Distribution
US and Canada
Product codes and lots
Part Number 6605632