Z-0444-05
Internal Battery for Puritan Bennett 700 Series Ventilator
Medtronic, Inc. / initiated 2004-12-16 / Terminated
Nellcor Puritan Bennett Inc began a recall of Internal Battery for Puritan Bennett 700 Series Ventilator on . The reason FDA records is "Wires on battery may loosen causing disconnection". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0444-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Internal Battery for Puritan Bennett 700 Series Ventilator.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Wires on battery may loosen causing disconnection.
- FDA root cause
- Other
- Action
- Letters will be Fedexed to customers which includes a simple test instruction that users can use to test the performance of the back up alarm.
- Quantity in commerce
- 6642 (3680 in vents and 2962 as replacements)
- Distribution
- Nationwide, Europe, Japan, Latin America and other portions of Asia.
- Product codes and lots
- All produced from 3/1/02 to 5/12/04