Nellcor Puritan Bennett Inc began a recall of Model 740 and 760 series Ventilators on . The reason FDA records is "Capacitor weakness in backup alarm system might fail causing lack of backup alarm to function in the event of total power loss". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0445-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Model 740 and 760 series Ventilators.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Capacitor weakness in backup alarm system might fail causing lack of backup alarm to function in the event of total power loss.
- FDA root cause
- Other
- Action
- Customers of record will be notified by letter sent FEDEX, a simple test instruction will be included to use to test the performance of the backup alarm.
- Quantity in commerce
- 13,010 worldwide, 3,948 in USA and territories.
- Distribution
- Worldwide.
- Product codes and lots
- All units ever manufactured. The product may be identified and tracked through serial numbers. This is a tracked device.