Z-0700-05
Product is CSN00210 8MHz blood flowmeter probe used with the Quantix/OR system
Neoprobe Corp. / initiated 2004-12-13 / Terminated
Neoprobe Corp began a recall of Product is CSN00210 8MHz blood flowmeter probe used with the Quantix/OR system on . The reason FDA records is "The probe's faceplate seal can fail and allow penetration of cleaning fluids during the cleaning and soaking cycle between uses or ingress of biohazard fluids during surgical procedures". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0700-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Neoprobe Corp.
- Product
- Product is CSN00210 8MHz blood flowmeter probe used with the Quantix/OR system.
- Product code
- DPW Flowmeter, Blood, Cardiovascular
- Reason for recall
- The probe's faceplate seal can fail and allow penetration of cleaning fluids during the cleaning and soaking cycle between uses or ingress of biohazard fluids during surgical procedures.
- FDA root cause
- Other
- Action
- The recalling firm notified cutomers by telepone and e mail beginning on 12/2/04.
- Quantity in commerce
- 64 devices
- Distribution
- Domestic accounts: Boston University Medical Center, 88 E. Newton St., B-402, Boston, MA 02118; Criticor, 2943 Pullman St., Santa Ana, CA 92705; Ilex Medical, 370 Cypress Creek Circle, Olsmar, FL 34677; Peerless Medical Corp., 705 Summer Oak Dr., Ellisville, MO 63021; Rolomac International, 1836 NE 213 Ln., Miami FL, 33179; University of Miami, PO Box 01690 (R-114), Miami, FL 33101. Foreign accounts: Asia Health Care Co Ltc., 4 FL-3, No 601, Chung Cheng Rd., Taipei, Taiwan, R.O.C.; Century Medical Inc., 1-6-4 Osaki, Shinagawa-KU, Tokyo, Japan 141-8588; Farmplex Cardic, Sheldare Enterprises, Ltd., 20-22 Bedford Row, London, United Kingdom; Integmed Ltd., 51 Hung To Road Kwun Tong, Kowloon, Hong Kong; Reysas Medical Corp., Hulya Sokal No. 21 GOP, 06700 Ankara, Turkey; Rahn AG, Dorflistrasse 12, CH-8050, Zurici, Switzerland.
- Product codes and lots
- All serial numbers.