Z-0453-05

OTC Pregnancy Test Kit labeled: ''CHOICE Pregnancy Test Kit'' (PA-320) , cassette type with cup and dropper...

International Chemical Inc. / initiated 2004-10-07 / Terminated / root cause: packaging

International Chemical Inc. began a recall of OTC Pregnancy Test Kit labeled: ''CHOICE Pregnancy Test Kit'' (PA-320) , cassette type with cup and dropper. Label states: ''Easy to Use, Easy Reading... on . The reason FDA records is "Poor sealing of pouches and degradation of product. Invalid test results and false negatives are possible". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0453-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
International Chemical Inc.
Product
OTC Pregnancy Test Kit labeled: ''CHOICE Pregnancy Test Kit'' (PA-320) , cassette type with cup and dropper. Label states: ''Easy to Use, Easy Reading, Positive result as early as 1 minute, Over 99% accurate.'' Each box contains one home test kit. Responsible firm on label: Distributed by: ICC, Amherst, NY 14228
Product code
LCX Kit, Test, Pregnancy, Hcg, Over The Counter
Reason for recall
Poor sealing of pouches and degradation of product. Invalid test results and false negatives are possible.
Root cause tag
packaging
FDA root cause
Other
Action
The consignee (exporter) was notified via telephone in October 2004 regarding lot U040904 (cassette) and notified via telephone regarding the expanded recall (lot U060804, One-Step) on 12/27/2004. A recall letter issued regarding both lots on 1/10/2005.
Quantity in commerce
108
Distribution
Product was shipped to an exporter in Florida who further distributed to consignees in the Caribbean.
Product codes and lots
U040904, Exp. 30SEP06; UPC code: 310960 451003