Z-0447-05

ATLANTIS Element Express Kit; a sterile kit for cervical spine surgery which includes hex-head screwdriver...

Warsaw Orthopedic, Inc. / initiated 2005-01-04 / Terminated

Warsaw Orthopedic, Inc. began a recall of ATLANTIS Element Express Kit; a sterile kit for cervical spine surgery which includes hex-head screwdriver, bone plate and 4 bone screws; catalog number 786-125 on . The reason FDA records is "The hex head of the screw driver may be too large to fit the screw heads of the bone screws included in the kit". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0447-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Warsaw Orthopedic, Inc.
Product
ATLANTIS Element Express Kit; a sterile kit for cervical spine surgery which includes hex-head screwdriver, bone plate and 4 bone screws; catalog number 786-125.
Product code
HWC Screw, Fixation, Bone
Reason for recall
The hex head of the screw driver may be too large to fit the screw heads of the bone screws included in the kit.
FDA root cause
Other
Action
Consignees were notified, via a recall letter dated 12/27/04 and issued on 1/4/05, to return the product. Competent Authorities were notified via letter dated 1/12/05.
Quantity in commerce
142
Distribution
Belgium, Croatia, Denmark, Finland, France, Hong Kong, India, Jamaica, Netherlands, Nicaragua, Portugal, Romania and Russia.
Product codes and lots
Lots 634550, 796050 and 796040.