Z-0533-05

Advisor Vital Signs Monitor model 9200, catalog number 92E654325. Item Description 9200 II...

Abbott Medical / initiated 2005-01-05 / Terminated

Smiths Medical PM, Inc. began a recall of Advisor Vital Signs Monitor model 9200, catalog number 92E654325. Item Description 9200 II 3LD/RP/O2/IP/TP/BATT ME. The basic monitoring package includes ECG... on . The reason FDA records is "An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0533-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Advisor Vital Signs Monitor model 9200, catalog number 92E654325. Item Description 9200 II 3LD/RP/O2/IP/TP/BATT ME. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
Product code
DPS Electrocardiograph
Reason for recall
An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
FDA root cause
Other
Action
On 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices.
Quantity in commerce
1
Distribution
Internationally only.
Product codes and lots
serial number 706460001