Z-0534-05
Advisor Vital Signs Monitor model 9200, catalog number 92E654335. Item Description 9200II...
Abbott Medical / initiated 2005-01-05 / Terminated
Smiths Medical PM, Inc. began a recall of Advisor Vital Signs Monitor model 9200, catalog number 92E654335. Item Description 9200II 3LD/RP/O2/IP/TP/PR/B ME. The basic monitoring package includes ECG (3... on . The reason FDA records is "An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0534-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Advisor Vital Signs Monitor model 9200, catalog number 92E654335. Item Description 9200II 3LD/RP/O2/IP/TP/PR/B ME. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
- Product code
- DPS Electrocardiograph
- Reason for recall
- An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
- FDA root cause
- Other
- Action
- On 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices.
- Quantity in commerce
- 7
- Distribution
- Internationally only.
- Product codes and lots
- serial numbers: 706462000, 706462002, 706462003, 706462005, 706462006, 706462007 and 706462009