Dental Explorer Single End #7, Mfr. Part Code/Product Code 100-7409. Distributed under the Henry Schein...
Hu-Friedy Mfg Co, Inc. began a recall of Dental Explorer Single End #7, Mfr. Part Code/Product Code 100-7409. Distributed under the Henry Schein label. Classification Name: 21CFR 872.4565; 510(k)... on . The reason FDA records is "Potential for tip breakage at the terminal shank due to improper heat-treatment at the foreign plant. Several reports of tip breakage were received by the own label distributor, and forwarded to the importer of record &...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0529-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- HU-Friedy Mfg. Co. LLC
- Product
- Dental Explorer Single End #7, Mfr. Part Code/Product Code 100-7409. Distributed under the Henry Schein label. Classification Name: 21CFR 872.4565; 510(k) exempt, Class I device, Device Listing # B057701. The product is labeled in part, HENRY SCHEIN 100-7409 Explorer - Single End #7 SE. +H65810074090C. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. --- This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling).
- Product code
- EKB Explorer, Operative
- Reason for recall
- Potential for tip breakage at the terminal shank due to improper heat-treatment at the foreign plant. Several reports of tip breakage were received by the own label distributor, and forwarded to the importer of record & specification setter of the dental explorers.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Hu-Friedy Mfg. Co., Inc. notified the Own Label Distributor of the recall on 1/7/05 by phone, and in writing on 1/10/05. Henry Schein, Inc. is having formal "URGENT: DEVICE RECALL" letters printed up which will be mailed to over 8100 of Schein's direct customers by First Class Mail on/about 1/21/05 to initiate the sub-recall of the dental explorers. Henry Schein will follow-up to non-responders by phone and/or fax.
- Quantity in commerce
- 42,517 total units distributed (all date codes)
- Distribution
- Dental practitioners, hospitals, dental clinics and distributors located nationwide, in Puerto Rico and internationally.
- Product codes and lots
- Date Codes/Lot Numbers: 0503, 0603, 0903, 1003, 1103, 1203. Barcode +H65810074090C.