Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the...
Diasorin Inc. / initiated 2004-04-20 / Terminated
Diasorin Inc. began a recall of Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay on . The reason FDA records is "An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0672-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay.
- Product code
- LLL Extractable Antinuclear Antibody, Antigen And Control
- Reason for recall
- An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.
- FDA root cause
- Other
- Action
- Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.
- Quantity in commerce
- 29
- Distribution
- CT, FL, NY, WA, TX, AR, TN, CANADA
- Product codes and lots
- software versions 9 thru 12