Z-0672-05

Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the...

Diasorin Inc. / initiated 2004-04-20 / Terminated

Diasorin Inc. began a recall of Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay on . The reason FDA records is "An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0672-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay.
Product code
LLL Extractable Antinuclear Antibody, Antigen And Control
Reason for recall
An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.
FDA root cause
Other
Action
Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.
Quantity in commerce
29
Distribution
CT, FL, NY, WA, TX, AR, TN, CANADA
Product codes and lots
software versions 9 thru 12