Z-0673-05

Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of...

Diasorin Inc. / initiated 2004-04-12 / Terminated

Diasorin Inc. began a recall of Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit on . The reason FDA records is "The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0673-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit
Product code
DBM Antimitochondrial Antibody, Indirect Immunofluorescent, Antigen, Control
Reason for recall
The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.
FDA root cause
Other
Action
A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.
Quantity in commerce
8
Distribution
CA, CT, KS, MA & WA
Product codes and lots
Software version 12