Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader)...
Diasorin Inc. began a recall of Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA... on . The reason FDA records is "Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0674-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.
- Product code
- MSV System,Test,Antibodies,B2 - Glycoprotein I (B2 - Gpi)
- Reason for recall
- Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.
- FDA root cause
- Other
- Action
- Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.
- Quantity in commerce
- 7
- Distribution
- CT, MI, MT, UT & CANADA
- Product codes and lots
- software version 12