Z-0674-05

Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader)...

Diasorin Inc. / initiated 2004-10-26 / Terminated

Diasorin Inc. began a recall of Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA... on . The reason FDA records is "Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0674-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.
Product code
MSV System,Test,Antibodies,B2 - Glycoprotein I (B2 - Gpi)
Reason for recall
Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.
FDA root cause
Other
Action
Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.
Quantity in commerce
7
Distribution
CT, MI, MT, UT & CANADA
Product codes and lots
software version 12