Z-0702-05

Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of...

Diasorin Inc. / initiated 2004-04-12 / Terminated / root cause: software

Diasorin Inc. began a recall of Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit on . The reason FDA records is "The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0702-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit
Product code
MLM Enzyme Immunoassay, Tracrolimus
Reason for recall
The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
Root cause tag
software
FDA root cause
Other
Action
Customers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.
Distribution
CA, Washington DC, FL, KY & Canada
Product codes and lots
version 12