Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the...
Diasorin Inc. began a recall of Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate... on . The reason FDA records is "When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0703-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
- Product code
- JTC Device, Microtiter Diluting/Dispensing
- Reason for recall
- When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Customers were notified by phone and informed of the problem and recommended users to use only the newer wash program which is identified is a 'W' in front of the program.
- Quantity in commerce
- 9
- Distribution
- CA, FL, NH, PA, SD, TX, WA
- Product codes and lots
- version 7