DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione...
Diasorin Inc. began a recall of DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione levels in serum. Catalog No. CA-1725. Lot # 04164... on . The reason FDA records is "Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0704-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione levels in serum. Catalog No. CA-1725. Lot # 04164. DiaSorin Stillwater, Minnesota 55082-0285, U.S.A.
- Product code
- CIZ Radioimmunoassay, Androstenedione
- Reason for recall
- Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used.
- FDA root cause
- Other
- Action
- Customers were sent notification letters informing them of the product and problem and recommended the destruction of any remaining inventory. Letter also requested the completion and return of the lower portion of the letter.
- Quantity in commerce
- 70 kits
- Distribution
- Massachusetts, Saudi Arabia, Sweden, Germany, Spain, Belgium, Italy, Camada. Australia, Portugal, Hungary, Israel,
- Product codes and lots
- catalog # CA 1725, lot # 04164