Z-0704-05

DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione...

Diasorin Inc. / initiated 2004-05-25 / Terminated

Diasorin Inc. began a recall of DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione levels in serum. Catalog No. CA-1725. Lot # 04164... on . The reason FDA records is "Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0704-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diasorin Inc.
Product
DiaSorin GammaCoat Androstenedione - 125 I, RIA Kit for the quantitative determination of androstenedione levels in serum. Catalog No. CA-1725. Lot # 04164. DiaSorin Stillwater, Minnesota 55082-0285, U.S.A.
Product code
CIZ Radioimmunoassay, Androstenedione
Reason for recall
Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used.
FDA root cause
Other
Action
Customers were sent notification letters informing them of the product and problem and recommended the destruction of any remaining inventory. Letter also requested the completion and return of the lower portion of the letter.
Quantity in commerce
70 kits
Distribution
Massachusetts, Saudi Arabia, Sweden, Germany, Spain, Belgium, Italy, Camada. Australia, Portugal, Hungary, Israel,
Product codes and lots
catalog # CA 1725, lot # 04164