PolyTiter software version 2.8 for use in the Polymedco PolyTiter Immunofluorescent Titration System...
Diasorin Inc. began a recall of PolyTiter software version 2.8 for use in the Polymedco PolyTiter Immunofluorescent Titration System, DiaSorin part # PT100 which is programmed for DiaSorin... on . The reason FDA records is "The program for a specific immunofluorescent titration system (Polymedco PolyTiter immunofluorescent Titration System) was found faulty in that the calibrator 2 value was incorrectly programmed as 1:120 instead of 1:160". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0676-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- PolyTiter software version 2.8 for use in the Polymedco PolyTiter Immunofluorescent Titration System, DiaSorin part # PT100 which is programmed for DiaSorin ANAFLUOR Indirect fluroescent Antibody Test For the detection and titration of circulating antinuclear antibodies. REF: 1624, 1662, 1660, 6624, 1604
- Product code
- DHN Antinuclear Antibody, Indirect Immunofluorescent, Antigen, Control
- Reason for recall
- The program for a specific immunofluorescent titration system (Polymedco PolyTiter immunofluorescent Titration System) was found faulty in that the calibrator 2 value was incorrectly programmed as 1:120 instead of 1:160.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- All customers were notified of the problem via phone and letter. Verbal instructions were given of how to change the Calibrator 2 titer to the correct one. A letter instructs customer to return software disc to DiaSorin, remove version 2.8 from the computer and install version 2.8.1.
- Quantity in commerce
- 14
- Distribution
- AL, AZ, CA, CO, FL, KY, MO, NC, NJ, NV, OH, TX, & VA
- Product codes and lots
- polyTiter software version 2.8