Z-0757-05
FlexTrayTM Endopath Gastric Tray, Kit Code TNG73
Ethicon Endo-Surgery Inc / initiated 2004-12-15 / Terminated
Ethicon Endo-Surgery began a recall of FlexTrayTM Endopath Gastric Tray, Kit Code TNG73 on . The reason FDA records is "Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0757-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- FlexTrayTM Endopath Gastric Tray, Kit Code TNG73.
- Product code
- MDM Instrument, Manual, Surgical, General Use
- Reason for recall
- Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.
- FDA root cause
- Other
- Action
- The recalling firm sent letters to hospitals and distributors, dated 12/14/04.
- Quantity in commerce
- 104 devices.
- Distribution
- The devices were distributed to hospitals and distributors located throughout the United States and to foreign consignees located in Saudi Arabia, Austria, Egypt, United Kingdom, Germany, France, Lebanon, The Netherlands, Spain, Pakistan, South Africa, Italy, United Arab Emirate, Czech Republic, Sweden, Switzerland, Greece, India, Portugal, Slovenia, Hungary, Palestine, and Kuwait.
- Product codes and lots
- V41T25, V42768, V42A65, V42F2D, V42H1U, V42N3Y.